For quantitative electro-chemical analysis in laboratories, Platinum Electrodes of various types are used. Generally, Electrodes are made of platinum 10% Iridium alloy for added mechanical strength which ensure long period of trouble free service. Electrode cylinders are usually made of wire gauze having wire diameter 0.15 mm * 360 mesh per cm2. This provides greater surface than area equal weight of plain or perforated platinum sheet and allows the electrolyte to circulate much more freely. Wherever agitation of electrolyte is required in order to save time of deposition, one of the rotating type of electrode is used.

For most electro-chemical analysis, wire gauze cathode is most satisfactory while a small size of similar anode is used if an anode deposit is required.

A simple helix of wire with stem is an adequate anode where there is no anodic deposit or when determining is not more than 0.2 percent lead as lead peroxide.

If desired, we can manufacture Platinum 10% Iridium electrodes to customers'own specifications to meet their special needs.

Care and Maintenance

A word of caution or care and maintenance of electrodes, more particularly wire gauze electrodes which are quite delicate compared to other Platinum ware.

 

  • Vigorous shaking should be avoided.

  • After electrolysis and weightments, the deposits are removed by dissolving in single acids without any damage to the electrodes.

  • Before re-use the electrodes should be cleaned with distilled water followed with drying around 100 C.

  • When determining Zinc, tin, lead, cadmium or mercury content by cathodic deposition, a prior electro deposit of copper or silver is recommended to avoid alloying with platinum.

  •  

Electrodes

Based on original design by Fischer, the following electrodes are used in stirred or unstirred electrolytes. The coil type (helical) anode is used when no anodic deposition is required or when determining less than 0.2 percent lead as lead peroxide.

Our Vision

* To embrace new technologies and methods. * To give unsurpassed products and services to the clients. * To constantly look for improvement and changes.

Site Designed & Maintained by Aman Infotech
Catalytic Gauzes, Glass Making Equipment, Pgm Solutions And Salts, Platinum Electrodes, Platinum Laboratory Equipments